Spectral Dynamics invented closed-loop digital vibration control and continues to lead with Panther — the next-generation control and analysis platform that combines precision, safety, and intuitive operation. Panther’s Random Control application builds on decades of engineering experience to deliver the industry’s most accurate, adaptive, and reliable random vibration control solution.

Panther Random Vibration Testing software with precise control and reliable test results. Analyze real-time random data or from recorded data.

How Random Vibration Test Works

Panther’s Random Control provides true Gaussian random generation with precise adaptive equalization for rapid loop closure and stable control across all spectral ranges. Engineers can configure narrowband or broadband profiles, overlay live and stored data, and monitor real-time response with full-resolution spectral plots. The advanced control architecture ensures superior amplitude accuracy and repeatable test results even under complex dynamic conditions.

Advanced Random-on-Random Control Technology

Click Icon for Full Screen or Esc to Return

Easy Setup and Flexible Test Configuration

Panther’s GTX environment simplifies random vibration testing through logical workflow and immediate access to critical data. Users can quickly configure tests, access stored configurations, and compare results using the Library interface. Touch-enabled controls and real-time graphical updates streamline operation for both new and experienced engineers.

Safety Features for Random Vibration Test

Panther’s embedded safety system continuously verifies control limits, channel integrity, and loop stability. Independent watchdog circuits and digital interlocks protect both the hardware and the device under test. Automatic aborts and adaptive gain management ensure that even under fault conditions, the test remains within safe limits.

Related Resources & Technical Documentation

Technical Data Sheet

Input Subsystem

Output Subsystem

Input Channels: 8 Channels Expandable up to 32

DAC: 20 Bit

Dynamic Range: >110dB

Max Output Amplitude: ±12 Volts Peak

ADC: 24 Bit

Max Output Current: 16 mA

Amplitude Accuracy: ±0.20% of value or ±0.03% of full scale

Amplitude Linearity: ±0.20% of value or ±0.03% of full scale

Voltage Ranges: ±12V to ±0.5V in 8 steps (Auto Ranging)

Max Output Rate: 262,144 Samples per Second

Overload Detection: Full scale on all channels, analog and digital detection

Image Attenuation: <96dB

Voltage Coupling: AC or DC or GND

Frequency Accuracy: ±5 ppm

IEPE Power: 4 mA (24V max into open circuit)

Freq Range Reduction: 10 stage smoothing filters

Max Rated Input Signal: ±35 Volts Peak

Reconstruction Filters: Analog 1kHz–25kHz, >96dB attenuation, ±0.15dB ripple

Max Sampling Rate: 262,144 Samples per Second

Output Impedance: 60 Ohms

Input Impedance: 1 MΩ

Output Type: Pseudo-Differential, 10Ω to Ground

Connection Type: Pseudo-Differential

Output Cable: Drives up to 50 feet of shielded 50Ω coax

Calibration: Automatic Internal Digital Calibration (NIST referenced)

Calibration: Internal Digital Calibration (NIST referenced)

TEDS Compliant: Yes (IEEE 1451.4)

Calibration Constants: Digital Calibration Constants stored in ASCII file

Host to Unit Communication: USB / Ethernet / Optional Wi-Fi

COLA Output Level: 1 Volt Peak (post-filter)